Emerson 3000 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-300NH Rev. 4.3
May 2005
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
World Class 3000
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
ADAPTOR PLATE NOT FURNISHED BY ROSEMOUNT
ANALYTICAL.
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
0.50 (13)
3.75 (95)
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
MTG HOLES
SHOWN ROTATED
45 OUT OF
TRUE POSITION
o
WELD OR BOLT ADAPTOR
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIR TIGHT.
0.50 (13)
4.50 (114)
O.D. REF
PIPE 4.00 SCHED 40
PIPE SLEEVE (NOT
BY ROSEMOUNT
)
LENGTH BY CUSTOMER
ANALYTICAL
MASONRY
STACK WALL
OUTSIDE WALL
SURFACE
JOINT MUST
BE AIRTIGHT
MTG HOLES
SHOWN ROTATED
45 OUT OF
TRUE POSITION
o
FIELD WELD
PIPE TO
ADAPTOR PLATE
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
NOTE: DIMENSIONS IN INCHES WITH
MILLIMETERS IN PARENTHESES.
2.50 (63.5)
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
WELD OR BOLT ADAPTOR
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIR TIGHT.
FIELD WELD
PIPE TO
ADAPTOR PLATE
3.50 (89)
O.D. REF
PIPE 3.00 SCHED 40
PIPE SLEEVE (NOT BY
ROSEMOUNT
) LENGTH
BY CUSTOMER
ANALYTICAL
MASONRY
STACK WALL
OUTSIDE WALL
SURFACE
JOINT MUST
BE AIRTIGHT
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
37840004
Figure 2-1. Probe Installation (Sheet 4 of 5)