Emerson 3000 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-300NH Rev. 4.3
May 2005
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-17
World Class 3000
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
-
-
-
BK
WH
J9
J8
J3
J2 J1
TRIAC RELAY
STACK
TC
ANALOG
HEATER
PROBE
MV
PROBE
MV
PROBE
TC
PROBE
TC
AD590
(INTERNAL
WIRING)
WHBK
OR CELL+
YE HTR TC +
BL CELL -
RD HTR TC
-
BK
WH
GN
STACK
TC
PROBE
HEATER
RH
N
L
LINE
VOLTAGE
GN/YE
TOP
BOTTOM
4 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
2 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
(OPTIONAL)
HPS 3000 INTERFACE MODULE
LINE
1
2
2
7
11
1
4
WORLD CLASS PROBE
ALWAYS DISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATION
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER
ON
120 V.A.C./100 V.A.C.
220/240 V.A.C.
JM4, JM1
JM5
REMOTE
ON
REMOVE JM2
INSTALL JM2
ELECTRONICS
SELECTION
JUMPER
PROBE HEATER
VOLTAGE SECTION
REMOVE JM3, JM6DIGITAL
(NEXT GENERATION)
WORLD CLASS PROBE JM7
NOTES:
12345678
BL
PROBE MV -
PROBE MV +
PROBE TC +
PROBE TC -
OR
YE
RD
GN
E
R
H
WH
BK
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
GN
BK
BK
}
HEATER
PROBE JUNCTION
BOX WIRING
A
B
37840007
RELAY WIRE IS OPTIONAL, RELAY CAN BE BYPASSED
WITH JUMPER JM-2 IF NOT WIRED TO THE IFT.
1
2
STACK TC WIRING AS REQUIRED.
3
ALL WIRES #16-#22 AWG TWISTED PAIR WITH SHIELD
BY CUSTOMER EXCEPT AS NOTED.
5
REMOVE JM1 ON INTERCONNECT BOARD.
6
INSTALL JM6 ON MICROPROCESSOR
BOARD PER FIGURE 2-8.
7
IF RELAY WIRE OF NOTE 1 INSTALLED THEN
REMOVE JM2 ON HPS 3000.
9
IF MPS 3000 NOT USED.
10
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION
STATUS INDICATION (48 V max, 100 mA max).
11
100 V.A.C. REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.
4
STANDARD PROBE CABLE BETWEEN PROBE AND
HPS BY ROSEMOUNT ANALYTICAL.
8
IF STACK TEMPERATURE NOT USED.
Figure 2-13. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 1of 2)