7. Place one (1) connector battery box onto battery sub-frame.
8. Slide one (1) connector battery box along the sub-frame until its guide pins are
engaged in the connector of the two (2) connector battery box.
NOTE: Visually inspect to ensure the connection is properly made. Connectors MUST be fully engaged.
CAUTION
The battery box retainer MUST be locked securely to hold the battery
boxes firmly in place or battery box connectors may be damaged causing
erratic wheelchair operation.
9. Pull the battery box retainer down over the end of the one (1) connector battery
box until it is securely clipped (locked) into place.
22NF BATTERY BOXES (FIGURE 10)
NOTE: To install the battery box onto the wheelchair, reverse the following procedure.
1. Place the wheelchair in a well ventilated area where work can be performed
without risking damage to carpeting or floor covering.
2. Verify the joystick ON/OFF switch is in the OFF position.
3. Disconnect the battery cable from the outside of the battery box.
4. Disconnect the battery box retention strap.
5. Remove the battery box.
6. Slide the four (4) clips that secure the battery box cover to the battery box to the
OPEN position.
NOTE: Arrows on the battery box cover point to the open position.
7. Remove battery box cover from the battery box.
FIGURE 10 - REMOVING/INSTALLING BATTERY BOXES -
22NF BATTERY BOXES
Battery Box
Retention Strap
Arrows
BATTERIESSECTION 9
BATTERIES