Emerson Rosemount 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4665, Rev AA
August 2010
Rosemount 8732
B-4
International Certifications
IECEx
E7 IECEx Flameproof Certificate No.: KEM 07.0038X
Ex de IIC T6 Gb or Ex de [ia Ga] IIC T6 Gb (-50 °C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max
= 250 V AC or 42 V DC
EF IECEx Flameproof Certificate No.: KEM 07.0038X
Ex de IIB T6 Gb or Ex de [ia IIC Ga] IIB T6 Gb (-50 °C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max
= 250 V AC or 42 V DC
NF IECEx Dust Certificate No.: KEM 07.0038X
Ex tD A20 IP66 T 100 °C
with I.S.: Ex tD A20 IP66 T 100 °C [Ex ia Ga] IIC
T6 (-50 °C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max
= 250 V AC or 42 V DC
Special conditions for safe use
(KEM 07.0038X):
Contact Rosemount Inc. for information on the dimensions of the flameproof
joints. The property class of the security screws which attach the flowtube or
junction box to the transmitter is SST A2-70, or SST A4-70.
Installation Instructions:
The cable and conduit entry devices and blanking elements shall be of a
certified flameproof or increased safety type, suitable for the conditions of use
and correctly installed. With the use of conduit, a certified stopping box shall
be provided immediately to the entrance of the enclosure.
N7 IECEx Type n Certificate No: IECEx BAS 07.0062X
Ex nA nL [ia] IIC T4 (-50 °C Ta +60 °C)
with LOI: (-20 °C Ta +60 °C)
V
max
= 42 V DC
Special conditions for safe use (x)
The apparatus is not capable of withstanding the 500V insulation test required
by Clause 6.8.1 of IEC 60079-15: 2005. This must be taken into account
when installing the apparatus.
NEPSI - China
E3 NEPSI Flameproof Certificate No.: GYJ071438X
Ex de IIC or Ex de [ia] IIC T6 (-50 °C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max
= 250 V AC or 42 V DC
EP NEPSI Flameproof Certificate No.: GYJ071438X
Ex de IIB or Ex de [ia] IIB T6 (-50 °C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max