Emerson Rosemount 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4665, Rev AA
August 2010
B-3
Rosemount 8732
European Certifications
E1 ATEX Flameproof
Certificate No.: KEMA 07ATEX0073 X II 2G
Ex de IIC or Ex de [ia] IIC T6 (-50°C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max
= 250 V AC or 42 V DC
0575
ED ATEX Flameproof
Certificate No.: KEMA 07ATEX0073 X II 2G
Ex de IIB or Ex de [ia] IIB T6 (-50°C Ta +60 °C)
with LOI (-20°C Ta +60 °C)
V
max
= 250 V AC or 42 V DC
0575
Special conditions for safe use
(KEMA 07ATEX0073 X):
Contact Rosemount Inc. for information on the dimensions of the flameproof
joints. The property class of the security screws which attach the flowtube or
junction box to the transmitter is SST A2-70, or SST A4-70.
Installation Instructions:
The cable and conduit entry devices and blanking elements shall be of a
certified flameproof or increased safety type, suitable for the conditions of use
and correctly installed. With the use of conduit, a certified stopping box shall
be provided immediately to the entrance of the enclosure.
N1 ATEX Type n Certificate No: BASEEFA 07ATEX0203X
II 3(1)G
Ex nA nL [ia] IIC T4 (-50 °C Ta 60 °C)
with LOI: (-20 °C Ta 60 °C)
V
max
= 42 V DC
0575
Special conditions for safe use (x)
The apparatus is not capable of withstanding the 500V insulation test required
by Clause 6.8.1 of EN 60079-15: 2005. This must be taken into account when
installing the apparatus.
ND
(1)
(1) Max surface temperature is 40 °C above the ambient temperature
conditions. Tmax = 100
°C
ATEX Dust Certificate No.: KEMA 06ATEX0006
II 1D
Ex tD A20 IP66 T 100 °C
with I.S.: [Ex ia] IIC II (1)G
Amb. Temp. Limits: (-20 °C Ta + 65 °C)
V
max
= 250 V AC or 42 V DC
IP 66