Rosemount 8732
EXTENDED DIAGNOSTIC
BYTES -
MANUFACTURER
SPECIFIC
MODULE DETAILS
Unit Diag Bit(34) Configuration invalid
Unit Diag Bit(35) Restart
Unit Diag Bit(36) Coldstart
Unit Diag Bit(37) Maintenance required
Unit Diag Bit(38) Characteristics invalid
Unit Diag Bit(39) Ident_Number violation
Unit Diag Bit(40) reserved (40)
Unit Diag Bit(41) reserved (41)
Unit Diag Bit(42) reserved (42)
Unit Diag Bit(43) reserved (43)
Unit Diag Bit(44) reserved (44)
Unit Diag Bit(45) reserved (45)
Unit Diag Bit(46) reserved (46)
Unit Diag Bit(47) reserved (47)
Unit Diag Bit(48) reserved (48)
Unit Diag Bit(49) reserved (49)
Unit Diag Bit(50) reserved (50)
Unit Diag Bit51) reserved (51)
Unit Diag Bit52) reserved (52
Unit Diag Bit(53) reserved (53
Unit Diag Bit(54) reserved (54)
Unit Diag Bit(55) Extension Available
Unit Diag Bit(16) Error appears
Unit Diag Bit(56) E01:DSP HW not cmptbl with SW
Unit Diag Bit(57) E02:Electronics Failure
Unit Diag Bit(58) E03:Coil Drive Open Circuit
Unit Diag Bit(59) E04:Empty Pipe Detected
Unit Diag Bit(60) E05:Electronics Trim Failure
Unit Diag Bit(61 E06:Auto Zero Failure
Unit Diag Bit(62) E07:Sensor Hi Limit Exceeded
Unit Diag Bit(63) E08:Sensor Processor Not Commn
Unit Diag Bit(64) E09:Universal Trim Failure
Unit Diag Bit(65) E10:Reverse Flow Detected
Unit Diag Bit(66) E11:Elex. Temp. Out of Range
Unit Diag Bit(67) E12:High Process Noise
Unit Diag Bit(68) E13:Grounding/Wiring Fault
Empty module Module EMPTY_MODULE 0x00 1 EndModule
Modules for flow meter Module AI 0x94 2 EndModule
Module TOTAL 0x41,0x84,0x85 3 EndModule
Module SETTOT_TOTAL"
0xC1,0x80,0x84,0x85
4 EndModule
Module SETTOT_MODETOT_TOTAL"
0xC1,0x81,0x84,0x85
5 EndModule