Emerson Rosemount 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4665, Rev AA
August 2010
Rosemount 8732
E-12
Model 10D1430 Sensor
(Integral) to
Rosemount 8732
Transmitter
Connect coil drive and electrode cables as shown in Figure E-9.
Figure E-9. Wiring Diagram for
Fischer and Porter Sensor
Model 10D1430 (Integral) and
Rosemount 8732
Table E-10. Fischer and Porter
Model 10D1430 (Integral)
Sensor Wiring Connections
Electrode Connections
ROSEMOUNT 8732 TRANSMITTER
To L2
1
2
3
7
6
8
L2
L1
U2
U1
1
2
3
7
6
L2
L1
U2
U1
G
TB1
TB2
Coil Connections
To Calibration Device
(Disconnect)
Rosemount 8732
Fischer and Porter Model 10D1430 (Integral)
Sensors
1 L1
2 L2
G
17 3
18 1
19 2
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor
or to terminals 1 and 2 of the
transmitter, or replacement of the
electronics board will be necessary.