Emerson Rosemount 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4665, Rev AA
August 2010
Rosemount 8732
6-8
To test the sensor, a multimeter capable of measuring conductance in
nanosiemens is preferred. Nanosiemens is the reciprocal of resistance.
or
Table 6-5. Sensor Test
Test
Sensor
Location
Required
Equipment
Measuring at
Connections
Expected Value Potential Cause Corrective Action
A. Sensor
Coil
Installed or
Uninstalled
Multimeter 1 and 2 = R
Open or
Shorted Coil
Remove and
replace sensor
B. Shields to
Case
Installed or
Uninstalled
Multimeter 17 and
and case
ground
17 and case
ground
Moisture in
terminal block
Leaky electrode
Process behind
liner
Clean terminal
block
Remove sensor
C. Coil Shield
to Coil
Installed or
Uninstalled
Multimeter 1 and
2 and
(< 1nS)
(< 1nS)
Process behind
liner
Leaky electrode
Moisture in
terminal block
Remove sensor
and dry
Clean terminal
block
Confirm with
sensor coil test
D. Electrode
Shield to
Electrode
Installed LCR (Set to
Resistance
and 120 Hz)
18 and 17 = R
1
19 and 17 = R
2
R
1
and R
2
should be stable
NO:
N5, E5, CD,
ED:
Unstable R
1
or
R
2
values
confirm coated
electrode
Shorted
electrode
Electrode not in
contact with
process
Empty Pipe
Low conductivity
Leaky electrode
Remove coating
from sensor wall
Use bulletnose
electrodes
Repeat
measurement
Pull sensor,
complete test in
Table 6-6 and
Table 6-7 on
page 6-10 out of
line.
2R18
0.2


R
1
R
2
300
R
1
R
2
1500
1nanosiemens
1
1gigaohm
----------------------------=
1nanosiemens
1
110
9
ohm
-------------------------------=