117 METER_VERIF_TEST_COND_IN Incoming Test Condition of the Flowmeter
Verification Test
Note: This parameter needs to be configured
before the Meter Verification is started via the
PERFORM_METER_VERIFY parameter.
118 METER_VERIF_TEST_COND_OUT Actual Test Condition of the Flowmeter
Verification Test.
119 METER_VERIF_CRITERIA Test limits that the Flowmeter Verification Test
was ran against.
120 METER_VERIF_RESULT Flowmeter Verification Test summary result.
121 COIL_RESIST_RESULT Coil Resistance Test Result, part of the
Flowmeter Verification Test.
122 COIL_INDUCT_RESULT Coil Inductance Test Result, part of the
Flowmeter Verification Test.
123 ELECT_RESIST_RESULT Electrode Resistance Test Result, part of the
Flowmeter Verification Test.
124 INT_SIM_RESULT Internal Simulator Test Result, part of the
Flowmeter Verification Test.
125 Reserved Reserved.
126 COIL_INDUCT_VALUE Coil Inductance value.
127 COIL_INDUCT_DEVIATION Coil Inductance Deviation.
128 COIL_INDUCT_FINGERPRINT Coil Inductance Fingerprint.
Note: These values should be made Read
Only in the DD. Only reset by the factory
parameter NV_RESET, not by the RESTART
w/ defaults.
129 COIL_RESIST_VALUE Coil Resistance value
130 COIL_RESIST_FINGERPRINT Coil Resistance Fingerprint
Note: These values should be made Read
Only in the DD. Only reset by the factory
parameter NV_RESET, not by the RESTART
w/ defaults.
131 ELECT_RESIST_VALUE Electrode Resistance value
132 ELECT_RESIST_FINGERPRINT Electrode Resistance Fingerprint
133 INT_SIM_DEVIATION The internal flow simulator reading as a
percent deviation from the reference value.
134 INT_SIM_REF_VALUE Internal Flow Simulator Reference Value
135 INT_SIM_VALUE The internal flow simulator reading.
136 METER_VERIF_EP_LIM Flowmeter Verification Limit – Empty Pipe
condition.
137 METER_VERIF_FLOWING_LIM Flowmeter Verification Limit – Flowing
condition.
138 METER_VERIF_NO_FLOW_LIM Flowmeter Verification Limit – No Flow
condition.
139 RECALL_FINGERPRINT_VALUES Perform a recall of the previous fingerprint
values.
140 PERFORM_REFINGERPRINT_FLO
WTUBE
Perform the re-fingerprint command.
Note: The parameter
FINGERPRINT_SELECT needs to be
configured before the Refingerprint operation
is started.
141 PERFORM_METER_VERIFY Perform the meter verification command.
Note: The parameter
METER_VERIF_TEST_SCOPE and
METER_VERIF_TEST_COND_IN needs to
be configured before the Refingerprint
operation is started.
Index Parameter Name Description