Rosemount 8732
A-4
Low Flow Cutoff
Adjustable between 0.01 and 38.37 ft./s (0.003 and 11.7 m/s). Below selected
value, output is driven to the zero flow rate signal level.
Overrange Capability
Signal output will remain linear until 110% of upper range value or 44 ft./s (13
m/s). The signal output will remain constant above these values. Out of range
message displayed on LOI and the Field Communicator.
Damping
Adjustable between 0 and 256 seconds
E-SERIES ADVANCED
DIAGNOSTICS
CAPABILITIES
Basic
Self test
Transmitter faults
Analog output test
Pulse output test
Tunable empty pipe
Reverse flow
Coil circuit fault
Electronics temperature
Process Diagnostics (D01)
Ground/wiring fault
High process noise
SMART Meter Verification (D02)
Smart Meter Verification
OUTPUT SIGNALS 8732 LOI Lockout
All optical switches on the display can be locked locally from the display
layout configuration screen by holding the upper right optical switch for 10
seconds. The display can be reactivated holding the same switch for 10
seconds.
PROFIBUS PA FIELDBUS
DIGITAL OUTPUT
SPECIFICATIONS
Output Signal
Manchester-encoded digital signal that conforms to IEC 1158-2 and ISA
50.02.
Profile Version
3.01
Identification Number
Generic: 0x9740
Manufacturer Specific: 0x0C15