8. Operation
Note:
When replacing objectives, you must perform a
teach-in for the new objectives in the Leica Ap-
plication Suite (LAS). A parfocality teach-in
should also be performed.
Note:
For lockable immersion objectives lock these by
pushing the front part upwards until it stops
(approx. 2 mm). Then, after a gentle turning mo-
tion to the right, the objective is locked.
For objectives with corrective mounts turn the
knurl to adjust the objective to the thickness of
the cover glass.
Color coding of objectives
The magnification of each objective is indicated by a color ring in accordance with DIN/ISO stand-
ards:
Immersion objectives are marked by an addi-
tional, lower color ring.
black oil or Imm (universal objective for
oil, water or glycerin)
white water
orange glycerin
The various engraved markings of the objectives
provide information on their applications:
black or bright field objectives,
dark blue strain-free
g
reen phase contrast objectives,
strain-free
100x 63x 40x 25x 16x 10x 6.3x 4x 2.5x 1.6x
125x 50x 32x 20x 5x
150x
160x
white dark- light- dark- light- yellow orange red brown gray
blue blue green green