Leica DMI4000B Microscope & Magnifier User Manual


 
94
8. Operation
8.7 Port selection
Leica DMI 3000 B:
Manual shifter rod activates and deactivates the
left-hand photo port.
Leica DMI 4000 B and Leica DMI 6000 B:
The
button on the front control panel switches 100%
of the light to the eyepieces.
Use the
button, also on the front control panel, to select
the side ports.
Depending on the configuration, the screen will
now display
- the active port (right or left) and
- the percentage of light going to the port
(100%, 80%, 50%).
Optional Leica DMI 6000 B:
The bottom port selection function can be as-
signed to one of the variable function buttons on
the stand or the SmartMove.
The top port can only be selected manually.
Select ports via
fixed function buttons on stand (side
ports)
variable function buttons on stand and
SmartMove (bottom port)
manual action (top port)
8.6 Tubes
Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Adjusting the viewing distance
Adjust the viewing distance of the eyepieces
so that a congruent total image is seen
(Fig. 115).
Adjusting the viewing angle
Ergotubes feature a tilting binocular section
for a 30–45° viewing angle adjustment range.
Beam splitting in photo tubes
The beam splitting is set manually by pulling out
a control bar.
Observation photo
VIS 100 % 0 %
100 % 0% 100 %
BL activation of Bertrand lens*
Light distribution via
manual control bar
Fig. 115 Tube
setting