Leica DMI4000B Microscope & Magnifier User Manual


 
63
7. Start-up
If a component has not been installed correctly,
the LeicaScreen will display an error message.
See Troubleshooting chapter, p. 102.
Components such as diaphragms, condensers,
light and phase rings have been pre-centered at
the factory. It may be necessary to correct the
centering after the microscope has been trans-
ported and assembled.
Before performing the required steps, please fa-
miliarize yourself with the LeicaScreen and the
controls.
Caution!
After turning on the gas discharge lamp, the
burner must be immediately adjusted. There-
fore, do not turn on the power supply unit
yet. First, work in transmitted light in order to
familiarize yourself with the microscope’s
controls.
Fig. 81
LeicaScreen
Initialization
7.2 Switching on the Microscope
Leica DMI3000B:
Switch on the microscope’s power at the On/
Off switch . The signal lamp is lit when the in-
strument is ready. (For the Leica DMI3000B
please continue at 7.4. Function Buttons on
the Stand)
Leica DMI4000B and Leica DMI6000B:
Switch on the power of the electronics box at
the On/Off switch (80.1). The signal lamp (80.2)
is lit green when the unit is ready. All motor-
ized microscope components will then run
through an initialization phase.
Note:
If a PC is connected, switch on the electronics
box first, and then the computer.
After the initialization (Fig. 81) is complete, the
LeicaScreen will display the microscope’s cur-
rent settings (Fig. 82).
Fig. 80
front side
Leica CTR6000
1 On/Off switch
2 Signal lamp
1
2
Fig. 82
LeicaScreen
after
Initialization