Leica DMI4000B Microscope & Magnifier User Manual


 
52
6. Assembly
6.13 Equipping the Incident Light Turret Disc
Caution:
Please read this section completely before be-
ginning with the assembly of the turret disk.
Leica DMI4000B and Leica DMI6000B:
The fluorescence drawer is located on the right
side of the stand. Before opening this drawer,
remove the cap below the drawer covering the
analyzer slot (66.1). Remove the analyzer if it is
already in the slot.
The replacement of individual cubes is more
convenient with the microscope switched on.
The position to be changed then automatically
turns to the outside and you can be sure that the
cube is positioned in the correct holder. You can
therefore postpone installing the filter cubes un-
til after the microscope has been switched on.
You can also insert the filter cubes while the in-
strument is switched off.
Press the white button next to the drawer. The
drawer will glide out into its initial position.
The inner disk can only be turned to accept the
fluo cubes in this position.
Fig. 66 Opening the fluorescence drawer
1 Analyzer slot
Fig. 64 Filter cube,
front side
Fig. 65 Filter cube,
back side
Fig. 67 Open fluorescence drawer
Fig. 68 Inserting or removing a filter cube
!
1