6. Assembly
6.12 Installation and replacement of
incident-light lamps
Caution!
Leica DMI4000B and Leica DMI6000B:
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
Ensure that the lamp housing has been dis-
connected from the power supply. Unplug
the power plug and the power supply during
assembly.
During assembly work on xenon burners, al-
ways wear the supplied protective gloves
and face protection (Fig. 58) (risk of explosion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path
(blinding hazard).
Lamp housing 106 z
This lamp housing is suitable for use with a 12 V
100 W halogen lamp or a variety of gas discharge
lamps.
Caution!
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow at least 30
mins for cooling down!
Fig. 58
Protective gloves and mask
Fig. 59 Lamp housing 106 z L with Hg 100 W lamp
1 Collector focusing
2 Vertical lamp adjustment
3 Horizontal lamp adjustment
4 Hg lamp mount
5 Reflector adjustment (not visible)
3
1
4