Leica DMI4000B Microscope & Magnifier User Manual


 
68
7. Start-up
Fig. 89 SmartMove remote control module
1 travel in x
2 Travel in y
3 Focus
4 Individual adjustment of button height
5 Variable function buttons (factory preset)
1
2
3
4
5
SET + Cube buttons 1-6
(Leica DMI4000B and DMI6000B only)
The SET command can be assigned to a function
button of the Leica DMI4000B.
The selected cube is moved to the loading posi-
tion for replacement. “Load” appears on the
screen. Press the opening button (88.2) to open
the drawer and position the appropriate cube.
The filter cube will remain in this position after
the drawer is closed.
7.5 The SmartMove Remote Control Module
SmartMove knobs
(Leica DMI4000B and Leica DMI6000B)
Use the knobs 89.1 and 89.2 to move the stage in
X and Y directions.
The image is focused using the knob 89.3 (Leica
DMI6000B only).
The height of the knobs can be adjusted to a
comfortable working position by turning 89.4.
Variable function buttons on SmartMove
The variable function buttons are assigned func-
tions at the factory that are appropriate to the
features of your microscope. They are labeled
accordingly. For details on button assignments,
please refer to the included identification sheet.
For information on the abbreviations used,
please refer to the list p. 62.
Note:
The Leica Application Suite (LAS) software is
required for changing the button assignments.
7.6 Illumination
7.6.1 Transmitted light
If your microscope has not yet been set up for
Koehler illumination, please continue with the
“Koehler Illumination” section.
Leica DMI3000B:
Select an objective with moderate magnifica-
tion
(10x–20x)
.
Set the condenser to the bright field position.
Place a specimen on the stage.
Focus on the specimen using the focus wheels.
Adjust the light intensity.
Close the field diaphragm manually until the
edge of the diaphragm appears in the field of
view.