Leica DMI4000B Microscope & Magnifier User Manual


 
50
6. Assembly
Caution!
Do not remove the burner’s dust cover until you
have installed the lamp. Avoid fingerprints on
the lamp. Sweat from your fingers on the glass
will shorten the life of the lamp significantly.
Install the burner in reverse order.
Caution!
Xe 75 burner:
Remove the burner’s dust cover (61.4) after
you have installed the burner.
Caution!
Make sure to follow the safety notes on
page 48.
•To open the 106 z lamp housing, unscrew the
fastening screws on the cover l. Loosen the
contact plug somewhat and pull it out of the
socket (63.9). Flip the cover up (63.1).
Loosen the mounting screws (63.8) on the
lamp socket and pull the socket out.
Remove the transport anchorage (red plastic
rod in place of the burner) in the lamp mount.
To do so, remove the lower clamp (60.1, 61.1).
Pull up the cooling element (61.3, 60.3) and
turn it to the side. Detach the lower clamp
system (61.2, 60.2) and remove the transport
anchorage.
Fig. 62 Rear panel of ebq 100 supply unit
1 Lamp connection
1
Fig. 63 106 z lamp housing (on the side, open)
1 Cover raised
2 Collector
3 12V 100W lamp or
gas discharge lamp in mount
4 Reflector (mirror)
5, 6, 7 Adjusting screw for x-y reflector
8 Locking screws for lamp mount
9 Socket for contact plug
1
2
3
898
4
5
6
7