Leica DMI4000B Microscope & Magnifier User Manual


 
86
8. Operation
8.2.4 Polarization (TL)
Leica DMI3000B:
Select an objective.
Set the condenser to the bright field position.
Remove all IC prisms from the light path.
Move the polarizer on the condenser into the
beam path.
Insert the analyzer into the right side of the
stand until it clicks into position.
Bring the polarizer and analyzer into cross po-
sition until they reach maximum darkness.
Insert a specimen.
Set the brightness at the light potentiometer
Focus the image with the focus wheels.
Leica DMI4000 B and Leica DMI6000 B:
Use the TL/IL function button to switch to
transmitted light (TL).
Select the POL (polarization) contrast method
by pressing the variable button POL.
Alternatively: press the variable button
CHANGE TL .
(For details on button assignments, please see
the identification sheet.)
POL will appear on the LeicaScreen.
Manual method:
Move the polarizer on the condenser into the
beam path.
Insert the analyzer into the right side of the
stand until it clicks into position (Fig. 110).
Bring the polarizer and analyzer into cross po-
sition until they reach maximum darkness.
Place a specimen on the stage and select a
suitable objective.
Motorized method:
If the microscope is equipped with the rel-
evant components, the polarizer will be acti-
vated automatically in the condenser when
the POL contrast method is selected. The
analyzer cube is also automatically positioned
in the beam path.
Combined methods:
The Leica DMI4000 B and Leica DMI6000B mi-
croscope permit purely mechanical and mo-
torized components – such as a mechanical
analyzer and motorized polarizer – to be com-
bined.
Fig. 110 Inserting the analyzer