Leica DMI4000B Microscope & Magnifier User Manual


 
87
8. Operation
8.2.5 Differential Interference Contrast (TL)
Leica DMI3000B:
Select an objective.
At the condenser, select the appropriate
Wollaston prism condenser.
At the objective turret, select the appropriate
Wollaston prism objective.
Move the polarizer on the condenser into the
beam path.
Insert the analyzer into the right side of the
stand until it clicks into position.
Insert a specimen.
Set the brightness at the light potentiometer
Focus the image with the focus wheels.
Use the knurled wheel below the objective
turret for fine adjustment (Fig. 111).
Leica DMI4000B and Leica DMI6000 B:
Use the TL/IL function button to switch to
transmitted light (TL).
Select the DIC contrast method
by pressing the variable button DIC.
Alternatively: press the variable button
CHANGE TL .
(For details on button assignments, please see
the identification sheet.)
DIC will appear on the LeicaScreen.
The polarizer in the condenser and the suit-
able condenser prism are automatically posi-
tioned in the beam path. The corresponding
objective prism and the analyzer cube are
also positioned automatically.
Place a DIC specimen on the stage.
Rotate an appropriate objective into place.
Focus the image with the knob on the
SmartMove or the focusing wheel and adjust
the intensity with the INT function buttons.
Use the knurled wheel below the objective
turret for fine adjustment (Fig. 111).
Manual alternative:
Move the polarizer on the condenser into the
beam path manually.
Insert the analyzer manually into the right side of
the stand until it clicks into position (Fig. 110).
Adjust the objective and condenser prisms
manually until a valid combination appears on
the display.
Use the knurled wheel below the objective
turret for fine adjustment (Fig. 111).
Fig. 111 DIC disk with knurled wheel for fine adjustment