What You See
on the Display
What You Do Comments
7-2
5. Scan or manually enter
the linearity kit lot
number via the keypad,
then press Enter. If the
monitor prompts for a
level to be entered, skip
to step 7.
If the linearity kit lot
number that was entered
is different from the
previous one, the monitor
will prompt the operator
to do one of two actions:
either re-enter the existing
linearity kit lot number or
replace the existing
linearity kit lot number
with the new linearity kit
lot number.
6. The operator can
perform one of the
following actions:
Press 1 – ReEnter
Kit Lot.
Press 2 – Replace Panel.
If 1 is pressed, the
monitor will return to the
previous screen so the
operator can scan or enter
the linearity kit lot
number.
The monitor will prompt
the operator to select a
level.
7. Enter the number of
the level of the next test
to be run.
There can be up to 4
replicates per level. The
number in the parenthesis
indicates the number of
replicates already run for
that level. When a level is
full, the menu item (1-
Level 1) will no longer
appear.
8. If the operator presses
6 for a new panel, the
monitor will again
prompt the operator to
select a level.
The monitor will
prompt the operator to
scan or manually enter
the test strip lot number.