Use a new test strip for each test.
Use only Precision PCx or Precision PCx Plus Test Strips.
If test results are out of range despite meeting the above criteria,
please repeat the test using a new box of control solutions and/or test
strips. If the results are still out of range, call Customer Care at
1-877-529-7185. Outside the United States, contact the local Abbott
Laboratories/MediSense office or distributor.
Troubleshooting: Out-of-Range Linearity Test Results
Repeat the test for that linearity solution to make sure that the operator
meets the conditions in this checklist:
No air bubbles are in the linearity bottle’s nozzle.
Calibrate the monitor using the barcode for the test strip used.
Enter the correct lot number for the linearity solutions.
Check storage temperature. For solution storage temperature range,
please refer to linearity kit instructions for use. Test Strips must be
stored between 39°F and 86°F (4°C and 30°C).
Check that the temperature and humidity conditions in the room
where the tests are being performed are within the acceptable
operating range. Refer to the Precision PCx or Precision PCx Plus
Test Strip and linearity kit instructions for use.
Check linearity kit insert for open bottle expiration date.
Use a new test strip for each test.
Use only Precision PCx or Precision PCx Plus Test Strips.
If test results are out of range despite meeting the above criteria, please
repeat the test using a new linearity kit and/or test strips. If the results
are still out of range, call Customer Care at 1-877-529-7185. Outside
the United States, contact your local Abbott Laboratories/MediSense
office or distributor.
Error Messages
In this section, you will find information relating to error messages
that appear on the display when errors are detected by the Precision
PCx Monitor.
For each corresponding message, an explanation is given and appropriate
responses are described. In many situations, it may be possible to proceed
with some of the functions, at least temporarily, before attending to the
problem. (For example, when the batteries are too low to permit testing, it
may be possible to review data for a short time).