MediSense PCx Blood Glucose Meter User Manual


 
Questions? Refer to Page 10-13 For Customer Care Information.
10-11
Symptom/Error
Message
Explanation How To Respond
During the test, the room
temperature became
unstable or moved outside
the limits within which
the monitor can perform
a test.
Ensure that the room
temperature is within the
specified limits. Allow the
monitor time to stabilize
to room temperature.
Please refer to the test
strip instructions for use
for appropriate operating
temperature range.
Press 1 to exit.
(Error codes 121, 122)
When a Control Test is selected:
A control solution lot
number has been entered
for a different level of
control test from the one
that the monitor expected
to run (in the usual low to
high sequence). The user
may choose to enter a
different lot number
(usually, for the expected
level) or to run the level
of test that matches the lot
number entered.
Press 1 to ReEnter the lot.
or
Press 2 to run the level
of test that matches the lot
number already entered.
This screen shows the
level of control test
the monitor expected
to run: Low, Normal,
or High, and the level
of the lot entered.
A new linearity kit lot
number has been entered,
different from the one in
memory. Linearity data
has not been uploaded
since the last test. (Only
one panel of data is
stored).
The kit lot number may
have been entered
incorrectly, or the user
may choose to replace
the earlier data.
Press 1 to enter a different
linearity kit lot number
(typically, the number of
the previous kit).
or
Press 2 to proceed, using
the new kit lot number.
The existing linearity data
will be replaced by data
from the new lot.
When a Linearity Test is Selected: