MediSense PCx Blood Glucose Meter User Manual


 
Apply a drop of blood directly to the target area on the test strip,
covering the entire area.
Note: Use VENOUS MODE only for venous samples that have been
in capped tubes and have not been mixed with air. Use CAPILLARY
or ARTERIAL MODE for all other samples.
How to Obtain an Arterial
Blood Sample:
Use only fresh whole
blood samples.
Clear the arterial line
before drawing a blood
sample into a syringe that
contains sodium heparin,
or lithium heparin
or EDTA.
Use the sample within
30 minutes of collection.
Mix the syringe several
times immediately before applying the sample to the target area
on the test strip.
Allow a drop of blood to form at the tip of the syringe.
Cover the entire target area on the test strip with the blood sample.
The syringe can briefly touch the test strip without affecting the
test result.
Disposing of Waste
Observe the following guidelines when disposing of
biohazardous waste:
Dispose of used lancets in an approved sharps container.
Discard used capillary tubes, disposable transfer pipettes or tips
and test strips in an approved biohazard container.
Follow your facility’s biohazard disposal policy.
Patient Test Procedure
Use the following procedure to perform patient tests.
Note: The word “Strip” on the display always refers to the test
strip. If the monitor displays a message not indicated in these
procedures, please refer to section 10, Troubleshooting.
Questions? Refer to Page 10-13 For Customer Care Information.
3-3
Precision PCx PlusTest Strip
with TrueMeasure Technology
For Use with Precision PCx
v2.2 Only
Precision PCx Test Strip
For Use with Precision PCx
v2.0, v2.1 and v2.2.