MediSense PCx Blood Glucose Meter User Manual


 
5-26
What You See
on the Display
What You Do Comments
Data Review for Linearity Data
The monitor stores the last linearity panel that includes a maximum
of 4 replicates for 5 levels (20 tests).*
Please refer to the beginning of the Data Review section for the step-
by-step instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was
selected, the following message will be displayed on the monitor:
“No Stored Tests”.
If the monitor displays a message not indicated in these procedures,
please refer to section 10, Troubleshooting.
1. Press 4 – Linearity
Data.
The monitor now shows the
most recent Linearity Test.
At the top of the display
the test level is indicated.
The Level indicates Level
1-5 while Test indicates
1-4 replicates at that level.
2. The operator can select
one of the following
options:
Press 1 – Previous, view
the previous test result.
Press 2 – Next, view the
next test result.
3. When finished
reviewing the data,
the operator can:
Press Clear to return to
the Data Review screen.
Press Menu to return to
the Menu Mode menu.
Press On/Off to turn off
the monitor.
The number that appears
to the right of 1-Previous
and 2-Next on the monitor
indicates the number of
tests available for review.
The display also shows
the following:
o: Operator ID
(15.3 mmol/L)
*It is outside the intended use of the PCx/QCM system to perform a test in linearity mode on a PCx
instrument and change the test result record to a patient result in QCM Data Management System.
Linearity test results outside the test strip assay range should only be used to set linearity
performance on the PCx instrument. The performance of the test strips has not been verified outside
the assay range listed on the test strip instructions for use.