MediSense PCx Blood Glucose Meter User Manual


 
The Linearity Test menu enables the operator to verify method linearity
using a Linearity Assessment Kit.*
Please use the following procedures to perform a Linearity Test.
If the monitor displays a message not indicated in these procedures,
please refer to section 10, Troubleshooting.
Questions? Refer to Page 10-13 For Customer Care Information.
1. Press On/Off to turn
on monitor.
The Abbott logo screen
will appear for a few
seconds, followed by the
product name, the
software version, then the
Test Mode menu.
2. Press Menu. The Menu Mode menu
will appear.
3. Press 3 – Linearity
Test.
The monitor will prompt
for the Operator ID to be
scanned or entered.
4. Press Scan or manually
enter the Operator ID
via the keypad, then
press Enter.
The monitor will next
prompt for the Linearity
Kit lot number to be
scanned or entered.
What You See
on the Display
What You Do Comments
*It is outside the intended use of the PCx/QCM system to perform a test in linearity mode on a PCx
instrument and change the test result record to a patient result in QCM Data Management System.
Linearity test results outside the test strip assay range should only be used to set linearity
performance on the PCx instrument. The performance of the test strips has not been verified outside
the assay range listed on the test strip instructions for use.
7-1
Linearity Test 7