Signal Output and Signal Input Parts
The following is recommended by UL for the “Intended signal input and
signal output connections complying with this requirement.” (UL 2601.1)
The use of ACCESSORY equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice shall
include:
Use of the accessory in the PATIENT VICINITY.
Evidence that the safety certification of the ACCESSORY has been
performed in accordance to the appropriate IEC 601-1 and/or IEC
601-1-1 harmonized national standard.
Patient Vicinity (UL 2601.1)
In areas in which patients are normally cared for, the patient vicinity is
the space with surfaces likely to be contacted by the patient or an
attendant who can touch the patient. This encloses a space within the
room 1.83 m (6 feet) beyond the perimeter of the bed (examination
table, dental chair, treatment booth, and the like) in its intended
location, and extending vertically 2.29, (7 1/2 feet) above the floor.
Equipment
Equipment shall be classified by marking and/or identification as
described:
1. Precision PCx Monitor – Class I (EQUIPMENT energized from an
external electrical power source).
2. Docking Station – Class II (EQUIPMENT energized from an
external electrical power source).
3. Rechargeable battery pack – Type B (degree of protection against
electric shock).