MediSense PCx Blood Glucose Meter User Manual


 
8-46
Change Setup for Edit Lists
The Precision PCx Monitor maintains two lists for the purpose of
system security.
1. The Operator List contains the ID numbers of operators who are
authorized to perform tests. When the operator enters an Operator
ID, the monitor checks the ID against the current list stored in the
monitor. The monitor can be set to one of the following options:
Allow Test No checking will occur.
Warn The monitor displays a warning message but allows
the operator to continue testing.
Lockout The monitor will not allow the operator to
continue testing.
The Change Setup for Edit Lists is used to make changes to the
Operator ID List. When a name is added to the list, an expiration date
(the date the operator's access to the monitor expires) must also be
entered. A single Operator ID or the whole list can be deleted or
changed. The Expiration Date for operators can also be changed.
2. The Strip Lot List contains the strip lot numbers of valid test strip
lots in the current inventory. When the operator enters a test strip lot
number (via the barcode scanner or by manually entering* the 14-
digit lot number), the monitor checks the number against the current
list stored in the monitor. The monitor can be set to one of the
following options:
Allow Test
No checking will occur.
Warn The monitor displays a warning message but allows
the operator to continue testing.
Lockout The monitor will not allow the operator to
continue testing.
The Change Setup for Edit Lists is used to make changes to the Strip
Lot List. The Strip Lot List can have test strip lots added or deleted.
The minimum and maximum control values for the test strip lots can
be customized to meet tighter default values needed by a facility.
If you have questions, please refer to page 10-13 for Customer
Care information.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.