MediSense PCx Blood Glucose Meter User Manual


 
10-8
Symptom/Error
Message
Explanation How To Respond
The Operator ID just
entered is not on the list
of acceptable Operator
IDs defined for this
monitor.
The menu item 2 –
Continue appears and
testing is permitted
because the Security:
Operator ID Not Certified
option is set to Warn.
1. Verify and re-enter the
ID using the barcode
scanner.
2. Manually enter the ID.
3. If the problem persists,
notify the system
administrator. Press 2
to continue.
The Operator ID entered
is not on the list of
acceptable Operator IDs
defined for this monitor.
Testing cannot continue
because the Security
Operator ID Not Certified
option is set to Lockout.
The Operator ID or test
strip lot number scanned
or entered has expired.
Testing cannot continue
because the Security
Operator ID Not Certified
Option is Set to Lockout.
The Operator ID entered
has expired.
2 – Continue appears and
testing is permitted
because Security Operator
ID Not Certified option is
set to Warn.
1. Verify and re-enter the
ID using the barcode
scanner.
2. Manually enter the ID.
3. If the problem persists,
notify the system
administrator.
Press 2 to continue with
testing, using the Operator
ID already entered.