MediSense PCx Blood Glucose Meter User Manual


 
10-6
Symptom/Error
Message
Explanation How To Respond
Less than 50 test data
records are available in
test memory for the
selected test type. The
Test Memory Full option
for that test type is set to
Prompt for Overwrite or
Lockout.
The number of test
records available is
displayed. Data should
be uploaded as soon as
possible.
For data upload:
1. To upload the data, do
one of the following:
• Place the monitor in the
docking station (if the
system has this option).
• Plug the communica-
tions cable into the data
port and connect to the
laptop. (Refer to QC
Manager manual).
2. If the problem persists,
notify the system
administrator.
Press 1 to continue
testing.
Test memory for the
selected test type is full.
The Test Memory Full
option for that test type
is set to Prompt for
Overwrite or Lockout.
(The menu item 2 –
Continue appears only
if this option is set to
Prompt for Overwrite.
Tests may be performed;
the oldest stored test
results of the selected type
will be overwritten with
new results).
If the option is set to
Lockout, data must be
uploaded before testing
can start. (Memory space
becomes available once
records have been
uploaded. They do not
have to be erased).
1. To upload the data, do
one of the following:
Place the monitor in the
docking station (if the
system has this option).
Plug the communications
cable into the data port
and connect to the laptop.
(Refer to QC Manager
manual).
2. If the problem persists,
notify the system
administrator.
Press 1 to exit to other
functions.
or
Press 2 to continue testing
and overwrite the oldest
stored results.
This screen appears
only when Patient
Test or Control Test
is selected.
This screen appears
only when Patient
Test or Control Test
is selected.