Leica DM IRB Microscope & Magnifier User Manual


 
44
Turn the IC turret (situated under the objective
nosepiece) until it clicks into the brightfield
position (H).
Operate the focus handwheel until the letter H
appears on the display panel as well. Turn the IC
turret by a quarter of a rotation into the next
clickstop area. The message “IC prism 2”
appears in the display. Read the marking on the
turret and set the electronic display to the same
code by turning the handwheel. Do this for all
four positions. Empty positions should be coded
“–”.
Installing the objectives
Select the “Objectives” option in the Learn
mode (OBJ) by pressing the “CHANGE” key; the
“OBJ” option now flashes.
Confirm your choice with the “LEARN” key.
Conclude the Learn mode for the IC turret by
pressing the “CHANGE” key. “EXIT” flashes in
the display; confirm with the “LEARN” key.
Ob j e c t i ve 1:
PH
O
E
X
I
T
5x
I C Pr i sm 1 :
H E
X
I
T
Learn mode: IC turret
Learn mode: Objective data – Magnification
Select objective “1” by pressing the objective
change keys (behind the focus handwheel on
the left).