Fig. 33 Lamphousing 106 z L
1 Lid, flipped up, 2 Collector, 3 12 V 100W halogen lamp or gas
discharge lamp in holder, 4, 9 Cover fixing screws, 5 Reflector,
6, 7, 8 x-y adjustment screw for reflector, 10 Fixing screws for
lamp mount, 11 Socket for contact plug
Lamphousing 106 z L
Slacken the fixing screw on the lid (33.10). Pull
the cut-out plug slightly out of the socket and flip
up lid (33.11; 33.1).
Move the collector to the front and lift the defect
lamp out of the base (33.1; 33.2; 33.3). For
convenience, the lamp holder can be removed
from the lamphousing as well. To do this,
slacken the fixing screws on the lamp holder
(33.10) and pull out lamp holder (Fig. 34).
Without removing its protective cover, put a new
lamp into the base, without tilting, as far as it will
go.
n. b.:
It is important to leave the protective cover
on the lamp until it is in position.
Avoid making finger marks or wipe off
immediately. Close the lamphousing.
1
2
3
4
5
6
7
8
9
10
10 11
Fig. 34 12 V 100 W lamp holder with halogen lamp