Fig. 90 Lamp holders for gas discharge lamps
1 Upper clamp, 2 Seal point of the burner, 3 Lower clamp, 4, 6 Drill holes for fixing the lamp holder, 5 Sockets for cut-out plug,
7 Protective cover
Hg 50
1
2
3
4
5
6
Xe 75
1
3
7
1
3
Hg 100
1
3
Hg 100
Stab.
Put the lamp holder with burner inserted into
the lamphousing and secure with the screws
(90.10).
Close the lid of the lamphousing. When
closing the lamphousing, make sure that the
pins of the cut-out plug engage in the sockets.
Retighten the screws of the lid.
Push the cut-out plug in as far as it will go.
Align the lamphousing against the microscope
and secure with the clamp screw.
Connect the lamphousing to the power unit
(compare mains voltage!).