The light slit diaphragm is now adjusted until the
bright stripe of the slit image is fully inside the
grey stripe of the modulator. The light slit
diaphragm can be rotated and moved in x and y
direction.
For the 10x objective the image of the modulator
and the light slit are virtually the same size.
Adjust the light slit diaphragm until the bright slit
lies near the dark edge.
Repeat this process for each objective.
Always make sure that the objective name and
the name of the light slit diaphragm coincide.
Then disengage the Bertrand lens with the
adjustment wheel on the right side of the
microscope. Switch on magnification 1x or
higher and put a specimen on the stage.