Leica DM IRB Microscope & Magnifier User Manual


 
34
n. b.:
Hold on to the tube adapter until the clamp
screw is tightened.
Then insert the HC FSA 25 PE tube in the change
mount of the tube adapter and fasten with clamp
screw.
The following tubes from the Leica DM R range
are also adaptable:
Bino HC BSA 25 (42.1)
Trino HC FSA 25 P and PR (42.2)
(P + PR = with and without back reflection)
Fig. 37 HCI B22, binocular tube with 45° viewing angle, field
of view up to 22 mm, eyepiece diameter 30 mm for HC PLAN
10x/20 or 22 eyepieces, interpupillary distance setting:
55 – 75 mm
1 Clamp screw, 2 Eyepiece port, 3 Siedentopf binocular part
Fig. 39 HCI BV22, ergo binocular tube with 15°– 50° viewing
angle, field of view index up to 22, eyepiece diameter 30 mm
for HC PLAN 10x/20 or 22 eyepieces, interpupillary distance
setting: 55 – 75 mm
1 Clamp screw, 2 Eyepiece port, 3 Siedentopf binocular part
Photo port for 1 photo/TV connection (43.2)
Photo port for 2 photo/TV connections (43.1)
Switchable 100 %/100 % (25.3)
All Leica DM R trinocular tubes have the
following beamsplitting system:
100 % vis., 100 % photo or 50 %/50 %.
!
1
2
3
1
5
2
4
3
1
3
2
Fig. 38 HCI 3T 22 trinocular tube with 45° viewing angle
Light path: 100% vis – switch rod
150% –50% – switch rod
100% –photo– switch rod
Field of view index up to 22, eyepiece diameter 30 mm for HC
PLAN 10x/20 or 22 eyepieces, interpupillary distance setting:
55 – 75 mm
1 Clamp screw, 2 Eyepiece port, 3 Siedentopf binocular part,
4 Photo/TV exit, 5 Switch rod