Philips 862478 Medical Alarms User Manual


 
Patient Safety Tests
4-42 Testing the Monitor
Enclosure
Leakage
Current
This test is compliance with EN60601-1 (Enclosure Leakage Current). Test at 110% of the
nominal line voltage.
The analyzer leakage current indication must not exceed the values listed below.
Table 4-5. Enclosure Leakage Current
EN60601-1 (1990 + A1, A2, A11, A12, A13) and UL2601 (2
nd
Ed. 1997) US Deviations
What to record on your service record:
S(3): P or S(3): F
Step Action
1 Connect the AC mains power cord to the analyzer as recommended by the
analyzer operating instructions.
2 Using the appropriate test cable, connect the analyzer to either of the screws on
the back of the monitor, next to the handle.
3 Turn the monitor on.
4 Perform the test as recommended by the analyzer operating instructions.
Test Condition Polarity Allowable Leakage Current
Normal Normal 100 µA
Reversed 100 µA
S.F.C
1
Open
Protective Earth
Normal 300 µA
Reversed 300 µA