Philips 862478 Medical Alarms User Manual


 
Visual Test
4-10 Testing the Monitor
Visual Test
Inspect the system for obvious signs of damage. Also check external leads and accessories.
What to record on the service record:
V:P or V:F
Power On Test
What to record on the service record:
PO:P or PO:F
.
Step Action
1 Switch the monitor On.
2 Observe whether the system boots up successfully without displaying any error
codes. The main monitoring screen should display.