Measuring and Displaying Parameters
12-10 Specifications
Trends
Parameter Specification
Type Graphical and tabular
Memory Storage 12 hours, nonvolatile
Data interval 20 seconds (stored data points is the
average over 20-second interval)
Graphical Trend
Format One graph per vital sign
Display range 2 hours, scrollable
Vertical scaling
Heart Rate 50-100, 0-250, 25-125 b/min
NBP 50-100, 0-150, 0-300 mmHg
SpO
2
0-100, 60-100, 80-100%
CO
2
0-100, 0-60, 20-50 mmHg
Respiration Rate 0-20, 0-50, 0-150 b/min
Temperature
o
C 15-45, 33-41, 35-39
o
C
Temperature
o
F 55-115, 91-107, 96-102
o
F
Tabular format One table for all variables.
Six fields per row (time and five vital
signs)
Display interval Per NBP measurement, or 15 minutes for
no NBP, or 20 seconds during alarm
condition