Philips 862478 Medical Alarms User Manual


 
Disassembly Procedures
Disassembly
7-15
Removing the
CO
2
Module
Note—This procedure is only for those that have the C3 Sedation model.
Step Action
1 Remove the 4 screws on the CO
2
module.
2 Tilt the CO2 module towards the NBP pump.
3 Gently pull the module away from the Main PCB module.
Note—
When reassembling the CO
2
module, ensure that the CO
2
header matches up with the mating header.
Screw #1 Screw #2
Screw #3
Screw #4