Philips 862478 Medical Alarms User Manual


 
Measuring and Displaying Parameters
Specifications
12-7
NBP (Non-
Invasive
Blood
Pressure)
Temperature
Parameter Specification
Standards Meets performance standards of ANSI/
AAMI SP10-1992
Technique Oscillometric
Cuff Pressure Display 10 to 300 mmHg
Pulse Rate Range 40 to 200 b/min
Blood Pressure Accuracy Mean error and standard deviation per
ANSI/AAMI SP10-1992
a
a. Systolic and Diastolic blood pressure measurements determined with this device are
equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation
method, and mean pressure measurements determined with this device are equivalent to
those obtained by an intra-arterial blood pressure measurement device, within the limits
prescribed by the American National Standard, Electronic or automated
sphygmomanometers.
Pulse Rate Accuracy Greater of ± 2 BMP or ± 2%
Initial Cuff Inflation 180 mmHg
Subsequence Cuff Inflation Prev SYS ±30 mmHg
Blood Pressure Measurement Range
Systolic 60 to 250 mmHg
Diastolic 40 to 220 mmHg
Mean Arterial Pressure 45 to 235 mmHg
Measurement Modes
Auto Automatic measurements at intervals of
1, 3, 5, 10, 30, 60 and 90 minutes
Manual Single measurement initiated by Start/
Stop button
STAT Series of consecutive measurements for
5 minutes
Parameter Specification
Technique Probe
Range 15 to 45
o
C (59 - 113
o
F)
Accuracy ± 0.1
o
C