Philips 862478 Medical Alarms User Manual


 
Diagnostic Mode
5-12 Configuring the Power-up Defaults Menu
CO
2
Test The CO
2
Test Screen is provided to facilitate troubleshooting problems and performing
verification testing to ensure the measuring results are accurate. See “CO2 Performance” on
page 4-29 for information on how to perform these tests.
WarningWarning
A FilterLine, connected to the monitor, should never be applied to a human being while
the monitor is in Diagnostic Mode. Injury could result.
DIAGNOSTIC MENU
CO2 TEST
CO2 Pressure in mmHg (ambient/cell): 745/740
Pump Voltage (Volts): 3.0
CO2 (mmHg): 0.00
Status:
Set Barometric Pressure (mmHg): 745
Flow Rate Check and Calibration:
CO2 Calibration: 07/16/2002 15:13
Pump Op Time: 58
Return