Philips 862478 Medical Alarms User Manual


 
Part 3 Using Support Functions
6-12 Troubleshooting
Error Code
Categories
As a reference, the following table lists the general categories for other error codes. The error
code categories are shown only in hexadecimal format.
64 The SpO
2
module is sending an
error message to the host CPU.
1. Ensure PicoSat module is
properly connected.
2. Replace the PicoSat module.
3. If problem persists, replace main
PCB module.
?The CO
2
module is sending an
error message to the host CPU.
1. Ensure Oridion module is
properly connected.
2. Rplace the Oridion module.
3. If problem persists, replace main
PCB module.
65, 66 PicoSat SpO
2
module has detected
an error during initialization.
1. Replace PicoSat module.
2. If problem persists, replace main
PCB module.
6E-71 PicoSat SpO
2
module has detected
an error on its serial port.
1. Replace PicoSat module.
2. If problem persists, replace main
PCB module.
Hex Code Explanation Recommended Action
Hexadecimal Code Explanation
500xxx internal user interface error
501xxx remove serial port error
502xxx date and time error
503xxx NBP error
504xxx front end error
505xxx alarm error
506xxx audio error
507xxx recorder error
508xxx trend error
509xxx flash memory data error
50axxx SpO
2
error
50bxxx ECG error
50cxxx power-down task error
50dxxx on-board diagnostic error
50exxx power monitor error
50fxxx temperature measurement error
510xxx internal user interface error
511xxx error handling error
513xxx serial driver error