Philips 862478 Medical Alarms User Manual


 
Test Map
Testing the Monitor
4-7
Test Map
The Test Map shows which tests are required in which situations.
Service Event
(When performing...)
Test Block Required
(...complete these test)
Installation Visual
Power On
Monitor exchange Visual
•Power On
Monitor opened Power On
Pneumatic Leakage Test
All Safety tests
Battery replacement Power On
Battery Performance
All Safety tests
Speaker replacement Power On
Pneumatic Leakage Test
All Safety tests
Backlight tube replacement Power On
Pneumatic Leakage Test
All Safety tests
NBP pump replacement Power On
Pneumatic Leakage Test
All Safety tests
Preventative Maintenance Tests
SpO
2
module replacement Power On
Pneumatic Leakage Test
All Safety tests
•SpO
2
Tests
CO
2
module replacement Power On
Pneumatic Leakage Test
All Safety tests
•CO
2
Tests
Power supply module replacement Power On
Pneumatic Leakage Test
Safety Tests
Patient monitoring I/O module
replacement
•Power On
Pneumatic Leakage Test
All Safety tests
All Performance tests
Communications module replacement Power On
Pneumatic Leakage Test
All Safety tests