Philips 862478 Medical Alarms User Manual


 
Disassembly Procedures
7-8 Disassembly
Front Case
Disassembly
After you have separated the front from the rear case you can replace the following from the
front case assembly:
Speaker
Backlight Tube
Caution
Before you remove any module, make sure you are wearing ESD protection and you are
working in a grounded environment.
5 Disconnect the display cable located on your left-hand side by pulling the
connector towards you. Disconnect the backlight tube, front panel and speaker
cables located on your right-hand side.
6 Separate the front from the rear case.
Step Action
Front Panel
Cables
Backlight Tube
Cables
Speaker Cable
Display Cable
Repeater