Test Reporting
Testing the Monitor
4-5
Test Reporting
The following table shows what must be recorded on the Service Record after completing the
tests in this chapter.
Where P = Pass, F = Fail and X is the measured value as defined in the tests described in this
chapter.
Test What to Record
Functionality Assurance Tests
Visual V:P or V:F
Power On PO:P or PO:F
Preventive Maintenance Tests
P NBP
PN:P/X1/X2/X3/X4/X5 or
PN:F/X1/X2/X3/X4/X5
See page 4-15 for details
Performance Procedures
CO
2
Performance PCO
2
: P/X1/X2/X3/X4/X5/X6/X7 or
PCO
2
: F/X1/X2/X3/X4/X5/X6/X7
See page 4-30 for details
Patient Safety Tests
Safety(1) Ground Integrity S(1):P or S(1):F
Safety(2) Earth Leakage Current S(2):P or S(2):F
Safety(3) Enclosure Leakage Current S(3):P or S(3):F
Safety(4) Patient Leakage Current S(4):P or S(4):F
Safety(5) Patient Leakage Current with Mains S(5):P or S(5):F