Diagnostic Mode
Configuring the Power-up Defaults Menu
5-7
Menu Item Function
Monitor On-Time Displays the number of hours, rounded
to the nearest hour, that the Main PCB
has been operational. This value
CANNOT be reset.
Backlight On-Time Displays the number of hours, rounded
to the nearest hour, that the LCD
Backlight has been operational. This
value can be reset to zero. An example is
when a technician changes the backlight
or installs a new LCD.
Recorder On-Time Displays the number of hours, rounded
to the nearest hour, that the C3 recorder
has been operational. This value can be
reset to zero. An example is when a
technician installs a new recorder.
Battery Deep Discharges Displays the number of deep-discharge
cycles seen by the battery. The monitor
records a deep discharge cycle when the
battery volt-age reaches 5.6 volts. At this
voltage, a Low Battery alarm is issued.
This value can be reset to zero. An
example is when a technician installs a
new battery.
System Software Version Displays the revision level of the system
software. The revision level is also
momentarily shown on the LCD as part
of the Copyright screen. The value
CANNOT be changed.
CO
2
Software Version Displays the revision level of the
software of the Oridion
®
CO
2
module.
This value CANNOT be changed.
SpO
2
Software Version Displays the revision level of the
software of the MP-205 SpO
2
module.
This value CANNOT be changed.
NBP Software Version Displays the revision level of the
software for the NBP module. This value
CANNOT be changed.
Return Exits the System Information Menu
and returns to the Diagnostic Menu.