Philips 862478 Medical Alarms User Manual


 
Disassembly Procedures
Disassembly
7-13
Removing the
Patient
Monitoring I/O
Module
Follow these instructions for removing the I/O module from the Main PCB module.
Step Action
1 Remove the following 2 screws to remove the connector module.
Note—
Use a flathead screwdriver to remove Screw #1.
2 Slide the I/O module out.
3 Disconnect the CO
2
connector and tubing and the NBP tubing. Slide the CO
2
exhaust tubing out.
4 Pull the I/O module away.
Screw #1
Screw #2
Patient
Monitoring
I/O Module