Philips 862478 Medical Alarms User Manual


 
Mounting
Site Preparations
2-7
Mounting
WarningWarning
Make sure that you have read all applicable instructions before attempting to install the
wall mount.
Wall mounts that are intended to support monitors must be capable of supporting 4
times the weight of the monitor when properly installed.
Do not mount any portion of a monitoring instrument over a patient’s bed.
Do not exceed the maximum rated load specified for each wall mount.
Ensure that no electrical wiring, utilities, or piping interfere with the selected wall
mounting location.
The rear panel includes threaded standoffs for attaching a mounting system accessory.