Philips 862478 Medical Alarms User Manual


 
Disassembly Procedures
7-4 Disassembly
Disassembly Procedures
This section provides instructions for closed case, front case and rear case disassembly
procedures.
Closed Case
Disassembly
Procedures
This section describes the items that can be removed without disassembling the main case of
the monitor.
Removing the
Battery
WarningWarning
Before you begin any disassembly procedure, you must remove the battery.
Step Action
1 With the rear of the monitor facing you, disconnect the battery cover from the
rear panel using a Philips no. 1 screwdriver to loosen the 4 captive screws
fastening the battery cover.
2 Slide the battery out by tilting the monitor towards you.