Fluke 1529 Chub-E4 Thermometer User Manual


 
The SENSe suffix, <chn>, specifies the channel (1 to 4). The parameter is
ohms for low range PRTs, Kohms for high range PRTs and thermistors, and
millivolts for thermocouples. The second returned parameter is always 0 for
PRTs and themistors and the reference junction temperature for thermocouples.
8.4.2 Measurement Control Commands
The measurement control commands are used to query and set the moving av
-
erage filter, the resistance range, the number of probe wires, and the sequence
timer. The measurement control commands for resistance range and the number
of probe wires are password protected.
8.4.2.1 INITiate
Provided for compatibility but has no effect.
INIT
8.4.2.2 INITiate:CONTinuous?
Provided for compatibility but has no effect.
INIT:CONT?
1
8.4.2.3 SENSe:AVERage:COUNt? [MIN|MAX|DEF]
Returns the moving average filter setting.
SENS:AVER:COUN?
4
Specifying the MIN parameter returns a value of 1, specifying the MAX pa
-
rameter returns a value of 10, and specifying the DEF parameter returns a value
of 1. The *RST command sets the average count to the default (1).
8.4.2.4 SENSe:AVERage:COUNt <num>|MIN|MAX|DEF
Sets the moving average filter setting.
SENS:AVER:COUN MIN
The <num> parameter must be a number between 1 and 10. Specifying the
MIN, MAX, or DEF parameter sets the moving average filter setting to the
minimum, maximum, or default, respectively. The *RST command sets the
moving average to 1.
8.4.2.5 SENSe<chn>:RESistance:RANGe? [MIN|MAX|DEF]
Returns the resistance range setting.
SENS3:RES:RANG?
91
8 Digital Communications Interface
Commands