Fluke 1529 Chub-E4 Thermometer User Manual


 
type. The acceptable conversion types depend on the channel type. The conver
-
sion types and their names are listed in Table 7 on page 128.
This command may be password-protected. To access this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protec
-
tion is disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-protec
-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com
-
mands).
8.4.4.5 CALCulate<chn>:CONVert:PARameter:CATalog?
Returns the names of the probe characterization parameters associated with the
current conversion type for the specified channel.
CALC2:CONV:PAR:CAT?
“RANGE”,”RTPW”,”A4”,”B4”,”A”,”B”,”C”,”D”
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). Each item in the response list is enclosed in double-quotes and
separated from other items by a comma. Each item indicates the name of a
characterization parameter for the current conversion type.If there are no pa-
rameters available for the current conversion type, an empty string is returned.
The list of parameters depends on the selected conversion type. Characteriza-
tion parameters for the various conversion types are listed in Table 8 on page
128, Conversion Types and Parameters.
8.4.4.6 CALCulate<chn>:CONVert:PARameter:VALue? [<param>|ALL]
Returns the value of the specified conversion parameter for the specified
channel.
CALC2:CONV:PAR:VAL? RTPW
100.0145
The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan
-
nel (5 to 14). The <param> parameter is the name of the parameter as given in
Table 8, Conversion Types and Parameters. If the <param> parameter does not
apply to the selected conversion type, an error -221, “Settings Conflict”, is
generated. Specifying the ALL parameter returns the name and value of all pa
-
rameters associated with the specified channel in a comma delimited list. If no
parameter is specified, the ALL parameter is assumed. The format of the re
-
sponse consists of a parameter name, enclosed in double quotes, followed by its
value (i.e. "param1",value1,"param2",value2,"param3",value3,...).
8.4.4.7 CALCulate<chn>:CONVert:PARameter:VALue <param>,<num>[,<
param>,<num>...]
Sets the value of one or more conversion parameters for the specified channel.
97
8 Digital Communications Interface
Commands